- Measured Gas: Oxygen (O2).
- Sensing Technology: Non-depleting paramagnetic percentage measurement.
- Area Classification: Safe area digital analyzer, wall-mounted IP66 enclosure for indoor and outdoor use.
- Measurement Range: 0–1 % to 0–100 % O2, lower detection limit < ±50 ppm O2.
- Sample Handling Options: High flow rate bypass cell, special transducer versions for solvent or hydrogen-bearing samples.
- Flow and Pressure Options: Optional flow alarm and pressure compensation for stable readings under changing conditions.
- Outputs: Isolated 4–20 mA analog output with configurable alarm and status relays.
- Digital Communications: Modbus RTU (RS485) and Modbus TCP (Ethernet) for integration and remote diagnostics.
- Dimensions and Weight: 448 mm (W) × 235 mm (H) × 227 mm (D); approx. 26 kg.