TC007 Process Turbidimeter

TC007 Process Turbidimeter

Robust Inline Turbidity Measurement with Infrared Backscatter and Attenuation Analysis.

The Kemtrak TC007 is a high-resolution, fiber optic-coupled NIR turbidimeter designed for real-time, in-line turbidity measurement across a broad range of concentrations. It utilizes a high-performance infrared LED light source and robust optics to combine attenuated and scattered light analysis using a proprietary nephelometric ratio algorithm. This ensures reliable, drift-free operation even in harsh industrial or hazardous environments. Automatic compensation for sample color and fouling ensures long-term stability, while a user-friendly interface enables configuration, calibration, and trending via HTML/Java WebUI.

Key Features

-
  • Real-time turbidity measurement from 0.01 to 1,000 FTU (extendable to 4,000 FTU)
  • Proprietary ratio algorithm combining 0°, 90°, and 180° light paths
  • LED light source with >100,000 h lifetime (850 nm TS AlGaAs IR)
  • Maintenance-free sapphire measurement cells with no moving parts
  • ISO 7027:1999(E) compliance for nephelometric measurements
  • Auto gain, auto zero, signal filtering, and 16 linearization tables
  • HTML/Java WebUI for configuration, validation, and trending
  • Configurable automatic cleaning and PID control capability

Applications

+
  • Filtration and centrifuge monitoring
  • Phase separation and interface detection
  • Leak detection and condensate carryover
  • Water in fuel / oil in water analysis
  • Quality control in food, pharma, chemical and paper industries

Technical Highlights

+
  • Measurement Method: Attenuation, scatter, and nephelometric ratio
  • Range: 0.01–1,000 FTU (extendable), 0–5 AU (attenuated), 0–100% solids (backscatter)
  • Resolution: Typically < ±0.05% of range
  • Accuracy: < ±2% of reading plus stray light | Repeatability: < ±1%
  • Stray Light: < 0.05 FTU (cell-dependent)
  • Display: 16×4 LCD + 5 status LEDs
  • Analog Output: 1–2 × 0/4–20 mA (0.025% resolution, <0.1% accuracy)
  • Relay Outputs: 4 × 1 A 240 VAC (alarm, PID, cleaning, failsafe)
  • Digital Inputs: 5 (product select, clean, zero, hold, etc.)
  • Remote Interface: RJ45 (HTML WebUI, MODBUS TCP)
  • Cleaning Control: Interval 15 min–2 months, duration 0–9999 s, auto zero/hold
  • PID Control: mA or pulse-width relay | Full gain/integral/derivative config
  • Data Logging: >17,000 points | Event Logging: >16,000 events
  • Housing: IP65 (analyzer), IP66 (flow cell) | Stainless steel AISI 304
  • Power Supply: 100–240 VAC, 50–60 Hz or 22–30 VAC/VDC | Max 25 VA

Ideal For

+
  • Pharma and biotech turbidity control
  • Food and beverage filtration monitoring
  • Chemical process phase and interface detection
  • Environmental and water analysis
  • Hazardous area in-line turbidity monitoring

Contact Us

Have a question or interested in a product? Feel free to reach out via the form, send us an email, or call us at +41 44 986 29 29 during business hours. 

"*" indicates required fields

Phone

+41 44 986 29 29

Email

welcome@muellerag.ch

Office location

Gewerbestrasse 9
CH-8132 Egg
Switzerland

"*" indicates required fields

This field is hidden when viewing the form
This field is hidden when viewing the form